HYG-2676电子式指针绝缘电阻表适于在各种电气设备的维修、试验及检定中作绝缘测试。有单电压机型、双电压机型和三电压机型,额定电压、量程合理配置成多种规格,适用面广。
1. 主要指标
型 号 |
2676 B |
2676 C |
2676 D |
2676 E |
2676 F |
2676 G |
|
2676 Bh |
2676 Ch |
2676 Dh |
2676 Eh |
2676 Fh |
2676 Gh |
||
额定电压(V) |
1000 |
2500 |
5000 |
5000 10000 |
2500 5000 |
500 1000 2500 |
|
测量电压 |
额定电压×(1±10%)负载阻值不低于量程的1% |
||||||
中值电阻 |
1GW |
1GW |
1GW |
1G 10GW |
1GW |
100MW 200MW 500MW |
|
测量范围 |
0~100GW 0~100MW |
0~200GW 0~200MW |
0~500GW 0~500MW |
0~1000GW 0~500GW |
0~500GW 0~250GW |
0~40GW 0~20GW 0~10GW |
|
准确度等级 |
5.0 |
||||||
误差为5% 的范围 |
(>.005~<20)GW (>.05~<2)MW |
(>.01~<50)GW (>1~<50)MW |
(>.02~<100)GW (>2~<100)MW |
(>.04~<200)GW (>.02~<100)GW |
(>.02~<100)GW (>.01~<50)GW |
>2MW~<10GW >1MW~<5GW >.5MW~<2.5GW |
|
误差为10% 的范围 |
(.002~.005)GW (20~100)GW (0.2~0.5)MW (20~100)MW |
(.005~.01)GW (50~200)GW (.5~1)MW (50~200)MW |
(.01~.02)GW (100~500)GW (1~2)MW (100~500)MW |
(.02~.04)GW (200~1000)GW (.01~.02)GW (100~500)GW |
(.01~.02)GW (100~500)GW (.005~.01)GW (50~250)GW |
(1~2)MW (10~40)GW (.5~1)MW (5~20)GW (.25~5)MW (2.5~10)GW |
|
输出短路 电流 |
≥ 2mA |
2. 刻度准确度等级的划分
表头有效刻度的起止点,为顶部带细横线(“T”)刻度点内的区间,分为Ⅰ、Ⅱ、Ⅲ区段,Ⅰ区段为起始刻度点到Ⅱ区段起始点(“ ”),Ⅲ区段为Ⅱ区段终点到最大有效量程点。
3. 计时器计时误差(仅对HYG2676※h型)
15s±0.2s; 60s±0.5s; 10min±5s
4. 其它指标
绝缘电阻: ≥50MΩ (1000V)
耐压: AC 5kV 50Hz l min
工作温度与湿度: -10℃ ~ +50℃ 85%RH
贮存温度与湿度: -15℃ ~ +55℃ 90%RH
电源: 8×1.5V(AA, LR6) 电池 或11.1V/1.8Ah锂电池
耗电: ≤280m A (E、L端短路时)
尺寸及重量:255mm (L)×135mm (W)×80mm (D) ≈1kg
适于在各种电气设备的维修、试验及检定中作绝缘测试。
有单电压机型、双电压机型和三电压机型,额定电压、量程合理配置成多种规格,适用面广。
容量大、抗干扰能力强。能满足高压、高阻、大容量负载测试的要求,示值准确、稳定、可靠。
绝缘阻值采用对数分度线,均匀清晰。双量程自动转换,读数方便。
带计时器(h型)仪表,可自动计时,并在15s、60s、10min各时刻锁定示值,发出声光提示,便于抄读示值。
低耗电,可选用8×1.5V(AA, LR6)碱性电池,
或11.1V/1.8Ah锂电池供电,
使用时间长。
具有电池容量检查功能。
具有防震、防潮、防尘结构,适应户外工作环境。
保护功能完善,能承受短路和被测电容残余电压冲击。